MDM Trouble Report for 2003 Oct 18 Telescope: 1.3m Observer(s): Marshall, Pepper (OSU) Instrument: Templeton Problem(s) Encountered: Same problem as last night. The filter wheel worked fine until 1:30 am, then it failed and timed out every time we tried to move it. We opened up the filter wheel and saw what the problem was: the filter wheel was spinning many times before finally reaching the selected position. We tried cycling the power to the MIS, restarting XMIS, pointing the telescope to zenith, but to no avail. Not sure what else to try... Not sure how to fix the problem, although I suspect one of the position sensors is going bad. Both nights the problem has happened when going from an I to a B filter more than halfway through the night. ------------------------------ Submitted on 2003 Oct 19 [5:20:11]